Thermocouple

Temperature Sensor Thermocouple Wire Handheld Probes for portable temperature indicators


THERMOCOUPLES & TEMPERATURE TRANSMITTER

Temperature Sensors

RTD (PT-100)
Class Ranges     Limits of Errors
A -200 to 650°C     ±0.15°C±0.2 %
B -200 to 850°C     ±0.30°C±0.5 %

Thermocouples




(Ranges & Limits of Errors as per ANSI)
Unless otherwise specified ‘Limits of Errors’ means ‘ Acceptable percentage of error on reading’. For the following table ‘Ranges II’ means ranges for which ‘Limits of errors’ are established.
DIN Mounting is also Available
TYPE COMPOSITION OF CONDUCTORS RANGES – I (°C) RANGES - II (°C) LIMITS OF ERRORS
B Platinum – 30% Rhodium/Platinum – 6% Rhodium 0 to 1820 871 to 1705 ±0.5%
E 10% Nickel Chromium/Copper – Nickel (Chromel Constantan) -270 to 1000 0 to 316 to 871 ±1.7°C
J Iron/Copper – Nickel (Ferrous Constantan) -210 to 1200 0 to 277 ±0.5%
±2.3°C
K 10% Nickel –Chromium/5% Nickel – Aluminium (Chromel/Alumel) -270 to 1372 277 to 760
0 to 277
±0.75%
±2.3°C
R Platinum/13% Rhodium – Platinum -50 to 1768 277 to 1260
0 to 538
±0.75%
±1.4°C
S Platinum/10% Rhodium – Platinum -50 to 1768 538 to 1482
0 to 538
±0.25%
±1.4°C
T Copper/Copper – Nickel (Copper Constantan) -270 to 400 538 to 1482
-101 to -59
-59 to 93
93 to 371
±0.25%
±2%
±0.9°C
±0.75%

MECHANISM
Two dissimilar electric conductors of specific different materials welded or soldered at one end (Hot junction), an Electro Motive Force (Thermo-Electric) is generated. This ‘EMF’ is proportional to the temprature difference between Reference point and the Measuring point.

DESIGNS
*Sheath (For corrosion, non corrosio, heat resistance etc.) Stainless Steel 304, 310, 321, 347, 446. Alloy 600, Ceramic, Brass, Incoloy, Inconel, Alumina and Other Materials suitable for specific applications.

Connection
Plain (non Mounting), Screwed, Flanged, Spring loaded.

Head
Cast Iron, Cast Aluminium, Corrosion Resistant Plastic, Flame Proof.

Insulation
Ceramic Bit, Mineral Insulation (MI)
(Mgo filled)

Wires
Simplex
Duplex
Triflex (T/C)

Grounded Type
The thermocouple wires are directly welded to the sheath

Exposed Type
Quick Response Bare thermocouple Wires beyond the sheath are welded to form a Junction.

Ungrounded Type
The thermocouple wires junction is electrically and mechanically insulated from the sheath.

HANDLE PROBES FOR PORTABLE TEMPERATURE INDICATORS.


IMMERSION PROBE
For general purpose.


NEEDLE PROBE (MI)
For quick response and small diameter.


AIR PROBE
For air and gas.


BOW CURVED SURFACE PROBE
For static convex and moving surfaces, rollers, pipes etc.


LEAF PROBE
For flat surfaces.


PRICK PROBE
For semi solids.